دو محصول دانش بنیان فاتحین صنعت شریف TEM و GTEM

دو محصول دانش بنیان فاتحین صنعت شریف سلولهای TEM و GTEM میباشد که جهت انجام تستهای انتشاری و ایمنی EMC استفاده میگردد. عمده تفاوت آندو بازة فرکانسی آنها میباشد که در TEM از فرکانس dc تا 3.5GHz پوشش داده میشود و اطلاعات تکمیلی آنرا در [ادامه] ملاحظه میفرمایید. اما بازه فرکانسی سلول GTEM از 30MHz الی 18GHz میباشد و اطلاعات تکمیلی آنرا در [ادامه] ملاحظه میفرمایید.

TEM

 

TEM CELL

Fatehin Sanat Sharif                                                                                                                                           www.sanat-sharif.com

 

TEM CELLTEM CELL

The transverse electromagnetic (TEM) cells are used for electromagnetic applications such as interference/compatibility (EMI/EMC) experiments of the small RF devices in a laboratory environment. The standard electromagnetic fields generated in the shielded environment of a TEM cell can also be used for calibration of the RF field probes. In this section, a 50 Ω open TEM cell is designed and fabricated to generate standard electromagnetic fields from DC up to 3.5GHz. To overcome the impedance mismatch and improve the voltage standing wave ratio (VSWR) along the TEM cell, a new piecewise linear tapering method is proposed for the inner conductor of the cell. We use some idea to improve reflection and transmission coefficient, especially transition coefficient. For example, slotted walls, slotted septum, resistors and long narrows ferrites. Our results are completely compatible with commercial instrument in the world and even in some parameters such as transmission coefficient are better than the others.

 

 

 

 

 

 

Specification

Input connector: N Nominal impedance: 50

Wave impedance: 377 Ohm

Nominal cell impedance: 50 Ohm

Frequency range: up to 3.5 GHz

Transmission loss: <1dB up to 1.5 GHz,

<2.6dB up to 2.5 GHz, <5dB up to 2.5 GHz

Input return loss: <-22dB up to 1.2 GHz,

<-15dB up to 2.2 GHz, <-10dB up to 3.5 GHz

Max. input power, W continuous/pulsed: 100W/200W

for time and frequency measurements

Outer (LxWxH) cm: 40x22x11

Septum Height: 50 mm

Weight Kg. Approx.: 20Kg

Max. test volume Rectangular area

(LxWxH)cm: 20x13x5

Application : field-immunity test, measuring of

emission and Antenna calibration

 

 